Current position:

Trace element analyzer


Trace element analyzer



The main parameters



Workbench


Mechatronics.


Lower detection limit


5ugCd (Ⅱ) / L; 0.01ugCd (Ⅱ) / L (dissolution method).


Precision


RSD≤1%, RSD≤5% (dissolution)


Anti-reduction ability


5000: 1


Detection resolution


⊿E≤35mV


Linear relationship


γ≥0.9990


Leakage current


Earth leakage current≤0.5mA; shell leakage current≤0.1mA.


Dielectric


Withstand voltage 4000V.


Detection range 1


0.1g hair: zinc-iron-magnesium-manganese; copper-lead-cadmium; calcium; phosphorus; aluminum; selenium.


Detection range 2


400ul serum: zinc; iron; calcium; magnesium; phosphorus; copper.


Detection range 3


40ul peripheral blood: zinc; iron; calcium; magnesium; phosphorus; copper.


Detection range 5


40ul peripheral blood: lead; cadmium.


Detection range 6


1ml urine: lead; cadmium.


Reagent


Zinc; iron; calcium; magnesium; copper; lead; cadmium. Original manufacturer reagents.


Reservoir


The test report can be stored, and the determination spectrum can be stored.


Statistical Analysis


Statistics can be made according to age, gender, occupation and other parameters.


Working parameter


Computer settings can be stored.


Scan speed


0.1mV / s ~ 5000V / s


sampling interval


0.0001 ~ 60000 s


Potential resolution


0.1 mV


Current sensitivity


≤50PA


automation


Microcomputer control, automatic sampling, configuration workstation.


Analytical method


Linear regression.


Workstation configuration


Industrial control computer, multimedia keyboard / optical mouse, liquid crystal display, color inkjet printer, full Chinese menu display, supporting mainstream operating systems such as Windows2000 and XP